4-20 mA Current Output Transmitter
for
AC RMS Voltage or Current Input
Responds
in 1 power line cycle response with 0.1% accuracy.
DIN Rail mounted, digitally
programmable.

Standard Features
- True AC RMS measurement in one 50/60 Hz
AC line cycle
- AC or DC coupling for signals from DC to 10
kHz
- 0.2, 2, 20, 200,
300V & 600V voltage ranges
- 2, 20, 200 mA and 5A
AC RMS current ranges
- 0.1% accuracy of full scale, 0% to 100%
of full scale
- All input ranges
selectable & factory calibrated
- Crest factor of 3.0 at
full scale
- 4-20 mA, 0-20 mA or 0-10V transmitter output,
jumper selectable, isolated
- Output resolution: 0.0015% of span
(16 bits)
- Output accuracy:
±0.15%
of span
- Easy setup with PC-based
Instrument Setup Software
- DIN rail mount housing only 22.5 mm wide
- Detachable screw-clamp connectors
- Dual solid state relays for alarm or
control, isolated
- RS232 or RS485 serial communications,
isolated
- Universal AC power, 85-264 Vac
- Priced at only $270
(basic AC transmitter)
Options:
- Extended main board
for custom curve linearization & rate from
successive readings
- Low voltage
power, 10-48 Vdc or 12-32 Vac
|

Description
The Laureate 4-20 mA output,
AC RMS input
transmitter provides six voltage input ranges and four current input ranges, all
factory calibrated and jumper selectable. A special 5.000A
range utilizes a built-in 0.01 ohm shunt to accept the output of 5A current transformers, with no need for a stepdown transformer. The current
readings can easily be scaled digitally as needed.
True RMS precision circuitry
is used to determine
the root-mean-square of complex waveforms from DC to 10 kHz. Spikes up to
3.0 times the
maximum of each range are accurately measured.
Accuracy
is 0.1% of full scale* for signals from DC to 10 kHz and
signal amplitude down to 0. The
crest factor (Vp / Vrms) is 3.0 at full scale, increasing to 300 for a
signal amplitude of 1% of full scale.
True RMS readings
in as little as one 50/60 Hz AC
line cycle are unmatched in the industry. Such a high read rate allows
anomalies to be detected and alarmed before they become expensive problems. On/off control and
alarm can be achieved dual solid state relays, which are standard.
Transmitter can also capture peak and valley for times as short as one 50/60 Hz AC power line cycle.
AC or DC coupling is jumper selectable. AC coupling
accommodates signals from 10 Hz to 10 kHz and is suitable for
applications such as measuring the ripple on a DC power supply. DC coupling
accommodates signals from 0 Hz to 10 kHz. Multiple integral cycles are
averaged for signals above 50/60 Hz. A single cycle is captured for signals
from 3 Hz to 50/60 Hz. Below 3 Hz and at DC, the capture rate is every 333
ms.
Use with current
Transformers. High common mode rejection allows stable
readings with current shunts located on the high side of the line. Five amp input
capability allows the output of 5A current transformers to be applied directly to the
transmitter, with no need for a stepdown transformer. The transmitter
output can easily be scaled
for the type of current transformer. Digital filtering is selectable for noisy signals.
Three analog transmitter
settings are
jumper selectable: 4-20 mA, 0-20 mA or 0-10V. The output tracks the reading
up to 99,999 counts
that would be transmitted digitally. For
AC RMS signals, this
reading has an accuracy of
0.1% of full scale from 1% to 100%, 10 Hz to 10 kHz, and can be
calibrated in software. Output isolation from signal and power grounds
eliminates potential ground loop problems. The supply can drive 20 mA into a
500 ohm (or lower) load for 10V compliance, or 10V into a 5K ohm (or higher)
load for 2 mA compliance.
Standard features of Laureate transmitters include:
-
Analog transmitter output, 16-bit, user scalable and isolated. User selectable 4-20 mA,
0-20 mA or 0-10V levels.
- Serial communications, isolated. User selectable RS232 or RS485, half or full duplex, Modbus or
Laurel ASCII protocol.
- Dual solid state relays for control or alarm, isolated.
Rated 130 mA at 140 Vac or 180 Vdc.
- Universal AC power supply for 85-264 Vac.
Easy Transmitter programming is via
Laurel's Instrument Setup Software,
which runs on a PC under MS Windows. This software can be downloaded from this website at no charge. The required transmitter-to-PC interface cable is available from
Laurel (P/N CBL04).
* For purposes
of accuracy calculation, the full scale range for the 300V and 600V scales
is 2000V (20,000 counts), and the full scale range for the 5A scale is 20A
(20,000 counts).
|

Specifications
Laureate
4-20 mA Transmitter
for
AC RMS Voltage or Current Input
|
AC RMS Input |
Range |
Resolution |
Input Ohms |
Error at 25°C |
|
Voltage Ranges
0% to 100% of FS,
10 Hz to 10 kHz |
200.00 mV
2.0000 V
20.000 V
200.00 V |
10 µV
100 µV
1 mV
10 mV |
1 MΩ |
± 0.1% of FS |
|
600.0 V |
100 mV |
1 MΩ |
± 0.3% of FS |
|
Current Ranges
0% to 100% of FS,
10 Hz to 10 kHz |
2.0000 mA
20.000 mA
200.00 mA |
0.1 µA
1 µA
10 µA |
100 Ω
10 Ω
1 Ω |
± 0.1% of FS |
|
5.000 A |
1 mA |
0.01 Ω |
± 0.4% of FS |
|
Input Resolution
A/D Conversion Ratae
Max applied voltage
Over-current protection
AC RMS
Crest Factor |
16
bits (65,536 steps)
50/sec at 50 Hz power, 60/sec at 60 Hz power
600 Vac for 20, 200, 600V ranges, 50 Vac other
voltage
ranges
25x for 2 mA, 8x for 20 mA, 2.5x for
200 mA, 1x for 5A
3.0x at 100% of full scale |
|
Analog Output (standard) |
|
Output Levels
Compliance, 4-20 mA
Compliance, 0-10V
Output Resolution
Output Accuracy
Output Isolation
Update Rate, > 50/60 Hz
Update Rate, 3 to 50/60 Hz
Update Rate, DC to 3 Hz |
4-20 mA, 0-20 mA, 0-10 Vdc (jumper selectable)
10V ( 0-500 ohm load )
2 mA ( 5 kOhm load or higher)
16 bits (65,536 steps)
0.02% of output span plus conversion accuracy
250V rms working, 2.3 kV rms per 1 minute test
60/s at 60 Hz power, 50/s at 50 Hz power
One signal cycle
Every 3.333 ms |
|
Dual Relay Output (standard) |
|
Relay Type
Load Rating |
Two solid state relays, SPST, normally open,
Form A
130 mA at 140 Vac or 180 Vdc |
|
Serial Communications (standard) |
|
Signal Types
Data Rates
Output Isolation
Serial Protocols
Modbus Modes
Modbus Compliance
Digital Addressing |
RS232 or RS485 (half or full duplex)
300, 600, 1200, 2400, 4800, 9600, 19200 baud
250V rms working, 2.3 kV rms per 1 min test
Modbus RTU, Modbus ASCII, Laurel ASCII
RTU or ASCII
Modbus over Serial Line Specification V1.0 (2002)
247 Modbus addresses.
Up to 32 devices on an RS485 line w/o a repeater. |
|
Power Input |
|
Standard Power
Low Power Option
Power Frequency
Power Isolation
Power Consumption |
85-264 Vac or 90-300 Vdc (DC operation not UL approved)
10-48 Vdc or 12-32 Vac
DC or 47-63 Hz
250V rms working, 2.3 kV rms per 1 min test
2 Watts |
|
Mechanical |
|
Dimensions
Mounting
Electrical Connections |
120 x 101 x 22.5 mm
35 mm rail per DIN EN 50022
Plug-in screw-clamp connectors |
|
Environmental |
|
Operating
Temperature
Storage Temperature
Relative Humidity
Cooling Required |
0°C
to 55°C
-40°C to 85°C
95% at 40°C, non-condensing
Mount transmitters with ventilation holes at top and bottom.
Leave 6 mm (1/4") between transmitters, or force air with a fan. |

Pinout
Laureate 4-20 mA Transmitter
for
AC RMS Voltage or Current Input


Ordering
Guide
Laureate 4-20 mA Transmitter
for
AC RMS Voltage or Current Input
Click on the radio buttons to build a model number with pricing (shown below the Ordering Guide).
|