Laureate™ Digital Stopwatch & TimerResolution to 0.2 µs for time measurement of single events
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
|
|
![]() |
![]() |
|
![]()
Standard Features
|
![]()
Description
Display. The event time (Item #1) may be displayed H, M or S format with six-digit resolution. The longest single-event timing interval is 999,999 hours. The highest resolution is 0.2 µs. The event time may also be displayed in HH.MM.SS clock format with 1 s resolution. The stopwatch display is updated during timing at a rate controlled by a gate time, up to 25/s. It is reset to zero when the next start pulse occurs. Accumulated time from multiple events (Item #2) is also tracked and may be displayed up to 999,999 hours.
Designed for flexibility. Optional plug-in boards for communications and control include single or dual isolated analog output boards, dual or quad relay boards, and serial communications boards. Laureates may be powered from 85-264 Vac or optionally from 12-32 Vac or 10-48 Vdc. The display is available with red or green LEDs. The 1/8 DIN case meets NEMA 4X (IP65) specifications from the front when panel mounted. Any setup functions and front panel keys can be locked out for simplified usage and security. |
![]()
|
Display |
|
|
Readout |
6 LED digits, 7-segment, 14.2 mm (.56"), red
or green |
|
Inputs |
|
|
Types |
AC,
pulses from NPN, PNP transistors, contact closures, |
|
Stopwatch Mode |
|
|
Timing Modes: |
|
|
Accuracy |
|
|
Time Base |
Crystal
calibrated to ±2 ppm |
|
Voltage, standard |
85-264 Vac or 90-300 Vdc (DC operation
not UL approved) |
|
Excitation Output (standard) |
|
|
5 Vdc |
5 Vdc ±5%, 100 mA |
|
Analog Output (optional) |
|
|
Output Levels |
4-20 mA, 0-20 mA, 0-10V, -10
to +10V (single-output option) |
|
Relay Outputs (optional) |
|
|
Relay
Types |
2 Form C
contact
relays or 4 Form A contact relays (NO) |
|
Serial Data I/O (optional) |
|
|
Board Selections |
RS232, RS485, Modbus RS485, USB, USB-to-RS485
converter |
|
Environmental |
|
|
Operating Temperature |
0°C
to 60°C |
|
Electrical Connections |
|
|
|
|
![]()
Applications
Stopwatch Mode |
|
![]() |
The stopwatch mode is used to time single events between start and stop pulses on the same channel. Duration of a single wave shape can be measured by tying the A and B channels together. |
Timing Process Dynamics |
|
![]() |
The start and stop pulses used for timing can be generated by the dual relay board in a Laureate analog panel meter or digital counter. For instance, the start and stop pulse edges can be created as temperature passes two alarm setpoints, or astemperature cycles in a hysteresis control mode. |
Replacing an Oscilloscope with a Laureate Meter |
|
|
An oscilloscope is great for viewing and timing pulses in a lab. However, in fixed installations where digital timing accuracy and control outputs are required, a low-cost Laureate time interval meter will be the instrument of choice. Resolution to 0.2 µs is feasible. |
| Instrumenting a Pulsed Laser System | |
![]() |
|
|
Some of the many possibilities in instrumenting a pulsed laser system with Laureate dual-channel counters: elapsed time, number of pulses, pulse width, pulse separation, duty cycle, and pulse rep rate. |
|
![]()
Ordering
Guide
Laureate™ Stopwatch Meters
Click on the radio buttons to build a model number with pricing (shown below the Ordering Guide).
|