Laureate™ Instrument Setup Software
FREE software to set up Laureate Digital Panel Meters,
Counters, Timers,
Remote
Displays,
and Transmitters.
 |
 |
 |
1/8 DIN Digital Panel Meter
|
DIN Rail Transmitter
|
1/8 DIN Counter, Timer
or Remote Display
|
Windows-based Instrument Setup (IS)
software is an easy
method to set up Laureate 1/8 DIN digital panel meters,
counters, timers and remote displays. It is the only
means to set up Laureate DIN rail transmitters. Laureate 1/8 DIN
instruments can also be set up from the front panel, as explained in their
respective Owners Manuals, but that
process is more difficult. Instrument Setup software is of benefit whether
or not the PC is connected to the instrument.
- When the PC is connected to the instrument,
Instrument Setup software can get the setup file from the instrument
or open a default setup file or previously saved setup file from
disk. View Setup then provides graphical user
interface (GUI) screens which pull-down menus applicable to input,
display, scaling, filtering, alarms, communications, analog output, and
front panel lockouts. Fields that are not applicable to the instrument
as configured are either left out or grayed out. Clicking on any item will bring up a detailed Help screen
for that item. Once edited, the setup file can be downloaded (or put)
into the instrument and be saved to disk. The same setup file can
then be downloaded into multiple instruments.
- When the PC is not
connected to the instrument, the above GUI screens can be used to
set up a virtual instrument. The setup file can then be saved to disk. Switching to
View Menu then brings up a screen with the
required front panel programming steps. This view can be printed out for
use at the instrument site and to serve as a hard copy record.
Instrument Setup software can be
downloaded at no charge from this website in the form of a
self-extracting 6 MB file named
instrument_setup.exe.
The latest version is 2_4_0. The same software can be used with Laureate digital panel meters, counter, timers, weight / scale meters, remote displays,
and transmitters.
|

RS-232 cable for 1/8 DIN instruments |
 |
For connection to a PC, each
1/8 DIN Laureate instrument must be equipped with a serial communications
board
and be connected to the computer via a
serial communications cable. The wiring of the RS-232 cable is illustrated
above with end views of the two connectors. This cable is available as P/N
CBL01. Plugging into the Laureate is
the RJ-11 connector of a common 4-wire telephone extension cable. Plugging
into the PC is custom-wired DB-9 to RJ-11 adapter, which accepts the other
end of the telephone extension cable. Following setup, the serial
communications board may be removed from the instrument if desired. All
Laureate transmitters come standard with a 3-wire serial interface, which
can be jumpered for RS232 or RS485.
To install, download
instrument_setup.exe
software from this website at no charge. Double-click on the downloaded
file to unzip into a special directory, such as c:\temp. Within
that directory, double-click on setup.exe, which will install the
software on your PC. You may wish to paste a shortcut (labeled IS2)
of the executable file Instrument Setup.exe onto the PC desktop.
|

Click on any of the reduced screens
below
for a full-size screen view, then click on the Back button of your
browser to return to this page. The screens examples below are for a
fully-loaded Series 2 Digital Panel Meter (DPM), which is connected to the PC via RS-232.
If the meter is
a Series 1 meter (pre-2007), this is sensed by the software, and somewhat different
screens are brought up. Please see
Series 1 setup screens.
|
 |
Welcome Screen
From the computer desktop,
click on Start > Programs > IS2 > IS2. Or click on the IS2
icon on your desktop. This splash screen will be displayed for three
seconds. The software revision number is in the lower right. |
 |
Communications Selection Screen
Specify your desired communication protocol and the serial
communications bus type, which should match the jumper setup of the
instrument. Select None if the PC is not connected to the
instrument. |
 |
Establish Communications Screen
If you selected RS-232, you will be asked to specify the PC Com Port
and Baud Rate, which should match the jumper setup of the instrument.
Click on Establish. With the right settings, the Communications
Established field will light up in green, and the Meter Type
will be recognized. If so, click on Main Menu. |
 |
Main Menu Screen
Click on File > Default Setup to retrieve the default setup
file from disk for your type of meter. Click on File > Open Setup
to retrieve a previously saved setup file from disk or on File > Save
Setup to save your edited setup file to disk. Click on DPM >
Get Setup to retrieve the setup file from your meter or on
DPM > Put Setup to download your edited setup file into the
meter. |
 |
DPM Input + Display Setup Screen
From the Main Menu, click on View > Setup, then on the
Input+Display tab. You can now specify the meter hardware, signal
type, display mode, and functions of control inputs A and B. Clicking on
any item brings up a pull-down menu with the available choices. |
 |
DPM Scaling Setup Screen
Click on the Scaling tab, which provides three scaling methods to
relate the signal to the displayed reading: 1) Scale and Offset method, 2)
Coordinates of two points method, and 3) Reading Coordinates of Two Points
method. The last method uses actual high and low signals, and the computer
will prompt you. |
 |
DPM Filter Setup Screen
Click on the Filter tab, which allows you to specify the digital
filter time constant (if any), the adaptive filter threshold, and whether
Peak
/
Valley values are filtered or unfiltered. As for all setup screens,
clicking on the F1 key while an item is highlighted brings up a Help
screen for that item, as illustrated. |
 |
DPM Relay Alarms
Setup Screen
Click on the Relay Alarms tab, which allows you to set
up Alarms 1 and 2 for the optional dual relay output board. Clicking on
any of the four numeric fields changes these to green and brings up a
special field to enter the desired numeric value, which is tied to the
displayed reading. |
 |
DPM Communications
Setup Screen
Click on the Communications tab so set up serial communications.
In particular, you can special the Serial Protocol and the meter address
if multiple meters are to be addressed on the same serial data line. |
 |
DPM Analog Output
Setup Screen
Click on the Analog Out tab so set up the optional analog output
board. Three output ranges are selectable, the endpoints of which can be
tied to user-specified High and Low readings. |
 |
DPM Lockouts Setup Screen
Click on the Lockouts tab to check off
menu items which will no longer be accessible from the front panel of the
meter. This will simplify meter operation and prevent unintended setup
changes. |

| Other Meter
Setup Utilities |
 |
DPM Front Panel Setup Screen
As an aid to programming the meter from the
front panel when a serial connection is not available, you can return to
the Main Menu and click on View > Menu. The required sequence
of front panel screens will then be displayed. Click on any step in the
sequence for the meaning of each digit, as illustrated for the FILtEr
step. For a hardcopy, simply press on Print. |
 |
DPM Jumper Setup Screen
From the Main Menu, click on Jumpers. A pull-down menu with all
plug-in board applicable to DPMs will be displayed. |
 |
DPM Jumper Setup Screens
Click on any of the displayed plug-in boards, and you will be presented
with the jumper positions and
electrical connections for your selected board. This minimizes the need to
refer to the printed manual. |
 |
DPM Commands Screen
From the Main Menu, click on Commands if your PC is connected to
the meter. You can then reset
individual meter functions via the PC, display current or peak readings, enter
numbers to be displayed remotely by the DPM, and more. |

 |
Graphical Output Screens
From the Main Menu, click on Readings
if your PC is connected to the meter. A pull-down menu then offers three
choices: List, Plot and Graph.
- List presents the latest readings
in a 20-row by 10-column table. Press Pause at any time to freeze the
display. This is one method to capture peak readings.
- Plot generates a plot of readings
vs. time in seconds. It effectively turns the DPM-PC combination into a
printing digital oscilloscope.
- Graph generates a histogram where
the horizontal axis is the reading and the vertical axis is the number
of occurrences of readings. The display continually resizes itself as
the number of readings increases.
|
 |
DPM
Calibration Screens
From the Main Menu, click on Calibration. A screen then allows
you to select each range to be calibrated, displays the required jumper
settings, and prompts you for the signals to be applied from an external
calibrator. |
 |
Frequency Meter Calibration
Screen
Calibration of the quartz
crystal of the Laureate frequency meter requires the input of a known frequency from a calibrator.
Apply the frequency, then enter the frequency in Hertz. Calibration will be
automatic, with storage of the calibration factor stored in non-volatile
memory. |
 |

|